Thoracic surgery
HOME > 상품정보 > Thoracic surgery
Oxygenator
CAPIOX® FX
  • Housing material Polycarbonate ​
  • Fiber material Microporous polypropylene ​
  • Heat exchanger material Stainless steel ​
  • Maximum pressure: Blood Inlet 133 kPa (1.000 mmHg)
  • Maximum pressure: Water Inlet 196 kPa (2 kgf/cm²)
Product Details

 

 

 

  

Description

Family of Hollow Fiber Oxygenators


A 32 µm screen filter surrounds the fiber layer of the oxygenator. Particulate micro-emboli that may be present in the blood are trapped in the filter mesh while gaseous emboli remain inside the oxygenator and in contact with the hollow fibers. Driven by the pressure difference, gaseous emboli enter the inner lumen of the microporous hollow fiber and are eliminated via the gas outlet.

 

 

Characteristics 



-Integrated arterial filter with self-venting technology

-Lower priming volume, high gas exchange and low pressure drop are optimally balanced for superb performance

-Hollow fibers manufactured exclusively by Terumo using a patented technology means total quality management from raw materials to finished product 

-Superior O2 transfer performance

 

General specifications

Oxygenator

Oxygenator

FX05

FX15

FX25

Housing material

Polycarbonate    

Fiber material

Microporous polypropylene    

Heat exchanger material

Stainless steel    

Fiber bundle surface area

Approx. 0.5 m²

Approx. 1.5 m²

Approx. 2.5 m²

Heat exchanger surface area

Approx. 0.035 m²

Approx. 0.14 m²

Approx. 0.2 m²

Blood flow range

Min. 0.1 L/min

 Min. 0.5 - 5.0 L/min

Min. 0.5 L/min

Max. 1.5 L/min

0.5 - 4.0 L/min*

Max. 7.0 L/min

Reference blood flow (AAMI std.)

2.5 L/min

7 L/min

N/A

Prime volume (static)

43 mL

144 mL

260 mL

Blood inlet port (from pump)

1/4" (6.4 mm)

3/8" (9.5 mm)

3/8" (9.5 mm)

Blood outlet port

1/4" (6.4 mm)

3/8" (9.5 mm)

3/8" (9.5 mm)

Cardioplegia Port

N/A

1/4" (6.4 mm)

1/4" (6.4 mm)

Luer port (for recirc. or blood cardioplegia)

One luer lock on blood outlet port

N/A  

Gas inlet port

1/4" (6.4 mm)    

Gas outlet port

5/16" (7.9 mm)

1/4" (6.4 mm)

1/4" (6.4 mm)

Water ports (Hansen quick-connect fittings)

1/2" (12.7 mm)    

Maximum pressure: Blood Inlet

133 kPa (1.000 mmHg)    

Maximum pressure: Water Inlet

196 kPa (2 kgf/cm²)    

* When using R30 hardshell reservoir 

 

 

 

Arterial filter 

 

Aterial filter

FX05

FX15

FX25

Pore size

32 µm    

Surface area

130 cm²

360 cm²

600 cm²

 

 

 

  

 

 Hardshell reservoir

 

 

hell reservoir

FX05

R40 (for FX15)

R30 (for FX15)

FX25

Housing material

Polycarbonate      

Venous pore size

 Pore size 47 µm       

Defoamer material

Polyurethane foam      

Blood flow range: Venous flow

Min. 0.1 L/min

Min. 0.5 L/min

Min. 0.5 L/min

Min. 0.5 L/min

Max. 1.5 L/min

Max. 5.0 L/min

Max. 4.0 L/min

Max. 7.0 L/min

Blood flow range: Cardiotomy inlet

N/A

Max. 5.0 L/min

Max. 4.0 L/min

Max. 5.0 L/min

Blood flow range: Combined flow 

N/A

Max. 4.0 L/min

Max. 5.0 L/min

Max. 7.0 L/min

Blood storage capacity

1,000 mL

4,000 mL

3,000 mL

4,000 mL

Minimum operating volume

15 mL

200 mL

70 mL

200 mL

Venous blood inlet port

1/4" (6.4 mm)
 rotatable

1/2" (12.7 mm)
rotatable

3/8" (9.5 mm) rotatable

1/2" (12.7 mm)
 rotatable

Blood outlet port (to pump)

1/4" (6.4 mm)

3/8" (9.5 mm)    

Suction ports

Five 3/16" - 1/4" (4.8 mm - 6.4 mm) rotatable

Six 1/4" (6,4 mm)    

Vertical port (to CR filter)

N/A

3/8" (9.5 mm)    

Quick prime port

1/4" (6.4 mm)